GC- - 9310 Gas Chromatograph

GC- - 9310 Gas Chromatograph

Network gas chromatograph can be linked with several chromatographs with one computer and one set of software to complete chromatographic data analysis, printing, storage, remote monitoring and remote transmission.  Equipped with 10M/100M self-adaptive Ethernet communication interface and...

Send Inquiry

Product Details

Network gas chromatograph can be linked with several

chromatographs with one computer and one set of software to

complete chromatographic data analysis, printing, storage,

remote monitoring and remote transmission.

 Equipped with 10M/100M self-adaptive Ethernet communication interface and

built-in IP protocol stack, remote data transmission function can be operated

through Intranet and Internet.

 3 independent connection modules: lab, supervisor A and supervisor B.

 Several gas chromatographs can be controlled by one set of NetChromTM

workstation to get /data processing and counter – control functions.

 With remote diagnosis and remote program update functions through

Internet (users permittance).

 128* 64 dots vacuum fluorescence display.

 Chinese and English version can be selected, suitable for WinXP, Win2000,

Win7, Win8 syste.

 The temperature control documents can be named and invoked by the user.

 Compacted with FID, TCD, ECD, FPD and NPD detectors, three detectors

an be installed simultaneously.

 Specialized modular structure design for repairing and replacement easily.

 Microcomputerized temperature control system with 6 chanels of

independent temperature controller and 16th-order programmed heating.

 Automatic rear door function.

 Electronic flow controller (EFC), electronic pressure controller (EPC)

functions.

 Timing power on function for on-line gas sample analysis(with on-line auto

sampler) .

 Functions of baseline storage and baseline deduction.

 Instrument can be connected with the automatic sampler produced by

other manufacturers, such as Hamamatsu model AOC-20I, Italia HTA

brand HT series automatic sampler, etc

Dimensions:560×480×480mm

Net Weight:50kg

AC Power Supply:220V±10%(50Hz±0.5 Hz)

Consume:≤2000W

Enviroment Temperature:+5℃~+35℃

Relative Humidity:≤85%

Column Box

Temperature

Control

Temperature Range  RT 5°C~399°C(Incremental 1°C)
Temperature AccuracyBetter than±0.1°C(200°C)

Temperature-programmed

magnification

10

Temperature-programmed

rate setting

0.1°C~40°C/min(Incremental 1°C)(1°C -200°C)

25°C/min(Incremental 1°C)(above 200°C)

Magnification temperature time0~655min(Incremental 1min)
Temperature objectcolumn box, detectors, samplers
TCDSensitivity≥3000mv.ml/mg
Baseline noise≤20μv
Baseline draft ≤30μv/30min
Current0~220mA(Incremental 1mA)
High sensitivity TCD≥5000mv.ml/mg
ECDDetection limit≤2×10 -13 g/s
Baseline draft≤2×10 -13 A/30min
Baseline noise≤1×10 -6 A
Linear range ≥103
FPDDetection limit≤5×10 -11 g/s, ≤1×10 -12 g/s (s element)
Baseline noise ≤5×10 -11 A
Baseline draft ≤5×10 -11 A/30min
FIDDetection limit ≤8×10 -12 g/s
Baseline noise ≤5×10 -14 A
Baseline draft≤1×10 -13 A/30min
Stabilization time30min

Temperature

(sampler,FID,TCD)

Temperature RangeRT 5°C~399°C(Incremental 1°C)
Temperature AccuracyBetter than ±0.1°C(200°C)
External Control1-3 chanels external controller0.01min (time control accuracy)
4th chanel external

0.01min (time control accuracy)

0.01min (close time,for external start singal only )

Install Requirements


Do not use in corrosive environment.

Table should be firm, and keep the distance between 30-40cm from the

surrounding objects, and do not place combustible materials.

Gas Requirements

DetectorGas purposeGas
Purity
FIDCarrier gasN 2 ≥99.999%
Carrier gasHe ≥99.999%
Combustion gasH 2≥99.999%

Combustion-

supporting gas

AirKeep dry
TCDCarrier gasH 2  ≥99.999%
Carrier gasHe ≥99.999%


Inquiry